High sensitive scheme for methane remote sensor based on tunable diode laser absorption spectroscopy |
Ding Wu-Wen, Sun Li-Qun, Yi Lu-Ying |
State Key Laboratory of Precision Measurement Technology and Instruments, Department of Precision Instruments, Tsinghua University, Beijing 100084, China |
|
|
Abstract Methane is an important raw material for various petrochemicals in industrial fields and as also a clean fuel in daily life. However, as an inflammable and explosive material, methane leak can lead to disastrous consequences such as fire and explosion. Furthermore, as a kind of greenhouse gas, methane has stronger influence on global warming than carbon dioxide. In this paper, we present a new high sensitive scheme for methane remote sensing, which can facilitate detection and location of methane leakage. And the 2v3 band (near 1653.7 nm) of methane is chosen as the target transition which is free from the absorption of the other molecule in atmosphere. A tunable distributed-feedback diode laser is adapted to scan across the target transition. A Fresnel lens with a diameter of 150 mm is employed to collect the ambient backscattering light from natural features such as the buildings. The first harmonic signal is used to normalize the second harmonic signal to remove the influence introduced by the unknown reflectance factor of the actual target, therefore no retro-reflector is needed. Traditional tunable diode laser absorption spectroscopy (TDLAS) method has difficulty in locating the second harmonic signal peak position in low concentration conditions because of low signal-noise-ratio (SNR). To improve the SNR especially in low concentration environment, a scheme named “baseline-offset” TDLAS is presented in the paper, in which a reference cell filled with standard methane sample is inserted into the measuring optical path. The reference cell can also be used to calibrate the sensor. Furthermore, the reference cell can be used to lock the central frequency of the diode laser to the absorption peak position to monitor concentration fluctuation continuously. In the peak-locking mode, the sensor demodulates the third harmonic signal as error signal to control the injection current of the laser source with PID control. Moreover, one advantage of peak-locking mode is that the measurement frequency is about two orders of magnitude higher than the traditional TDLAS method. With “baseline-offset” TDLAS, the remote sensor described in this paper obtains SNRs as high as 19 and 16 at a stand-off distance of 10 m and 20 m, respectively. With such a high SNR, there is no necessity for complex algorithm in absorption peak position location. By defining the standard deviation of the measuring concentration as the detection limit, experimental results show that the proposed methane remote sensor has detection limits of 5 ppm · m at a distance of 10 m and 16 ppm·m for 20 m, respectively, while measuring the ambient methane. In peak-locked mode, the experimental system has a detection limit of 22 ppm·m at a distance up to 37 m and can monitor rapid concentration fluctuation in.
|
Received: 06 January 2017
|
PACS: |
07.07.Df
|
(Sensors (chemical, optical, electrical, movement, gas, etc.); remote sensing)
|
|
42.62.-b
|
(Laser applications)
|
|
42.62.Fi
|
(Laser spectroscopy)
|
|
42.68.Ca
|
(Spectral absorption by atmospheric gases)
|
|
Fund: Project supported by the National Major Scientific Instrument and Equipment Development Project of China (Grant Nos. 2012YQ200182, 2012YQ0901670602). |
Corresponding Authors:
孙利群
E-mail: sunlq@mail.tsinghua.edu.cn
|
|
|
|
[1] |
Fukada S, Nakamura N, Monden J 2004 Int. J. Hydrogen Energ. 29 619
|
[2] |
Fincke J R, Anderson R P, Hyde T, Detering B A, Wright R, Bewley R L, Haggard D C, Swank W D 2002 Plasma Chem. Plasma P. 22 105
|
[3] |
Mer J L, Roger P 2001 Eur. J. Soil. Biol. 37 25
|
[4] |
Iseki T, Tai H, Kimura K 2000 Meas. Sci. Technol. 11 594
|
[5] |
Zhang S, Liu W Q, Zhang Y J, Ruan J, Kan R F, You K, Yu D Q, Dong J T, Han X L 2012 Acta Phys. Sin. 61 050701 (in Chinese) [张帅, 刘文清, 张玉钧, 阮俊, 阚瑞峰, 尤坤, 于殿强, 董金婷, 韩小磊 2012 物理学报 61 050701]
|
[6] |
Wainner R, Green B, Allen M G, White M, Stafford-Evans J, Naper R 2002 Appl. Phys. B 75 249
|
[7] |
Goldenstein C S, Mitchell Spearrin R, Hanson R K 2016 Appl. Opt. 55 479
|
[8] |
Kan R F, Liu W Q, Zhang Y J, Liu J G, Dong F Z, Gao S H, Wang M, Chen J 2005 Acta Phys. Sin. 54 1927 (in Chinese) [阚瑞峰, 刘文清, 张玉钧, 刘建国, 董凤忠, 高山虎, 王敏, 陈军 2005 物理学报 54 1927]
|
[9] |
Xia H H, Kan R F, Liu J G, Xu Z Y, He Y B 2016 Chin. Phys. B 25 064205
|
[10] |
Rieker G B, Jeffries J B, Hanson R K, Mathur T, Gruber M R, Carter C D 2009 Proc. Combst. Inst. 32 831
|
[11] |
Huang Q B, Xu X M, Li C J, Ding Y P, Cao C, Yin L Z, Ding J F 2016 Chin. Phys. B 25 114202
|
[12] |
Chakraborty A L, Ruxton K, Johnstone W, Lengden M, Duffin K 2009 Opt. Express 17 9602
|
[13] |
Nadezhdinskii A, Berezin A, Chernin S, Ershov O, Kutnyak V 1999 Spectrochim. Acta A 55 2083
|
[14] |
Reid J, Labrie D 1981 Appl. Phys. B 26 203
|
[15] |
Duffin K, McGettrick A J, Johnstone W, Stewart G, Moodie D G 2007 J. Lightwave Technol. 25 3114
|
[16] |
Fernholz T, Teichert H, Ebert V 2002 Appl. Phys. B 75 229
|
[17] |
Cao Y N, Wang G S, Tan T, Wang L, Mei J X, Cai T D, Gao X M 2016 Acta Phys. Sin. 65 084202 (in Chinese) [曹亚南, 王贵师, 谈图, 汪磊, 梅教旭, 蔡廷栋, 高晓明 2016 物理学报 65 084202]
|
[18] |
Kluczynski P, Axner O 1999 Appl. Opt. 38 5803
|
[19] |
Rothman L S, Gordon I E, Babikov Y 2013 J. Quant. Spectrosc. Ra. 130 4
|
[20] |
Werle P W, Mazzinghi P, D'Amato F, Rosa M D, Maurer K, Slemr F 2004 Spectrochim. Acta A 60 1685
|
[1] |
Ding Wu-Wen, Sun Li-Qun. Phase sensitive chirped laser dispersion spectroscopy under high absorbance conditions[J]. Acta Physica Sinica, 2017, 66(12):
.
doi:10.7498/aps.66.120601. |
[2] |
Feng Ming-Chun, Xu Liang, Liu Wen-Qing, Liu Jian-Guo, Gao Min-Guang, Wei Xiu-Li. Investigation of detecting biological aerosol by passive Fourier transform infrared spectroscopy technology based on MODTRAN model[J]. Acta Physica Sinica, 2016, 65(1):
.
doi:10.7498/aps.65.014210. |
[3] |
Zhang Shu-Feng, Lan Li-Juan, Ding Yan-Jun, Jia Jun-Wei, Peng Zhi-Min. Theoretical and applied researches on measuring line width in wavelength modulation spectroscopy[J]. Acta Phys. Sin, 2015, 64(5):
.
doi:10.7498/aps.64.053301. |
[4] |
Liu Yong-Jun, Wang Fei-Ru, Sun Wei-Min, Liu Xiao-Qi, Zhang Ling-Li. Tunable capability of dye-doped cholesteric liquid crystal lasers[J]. Acta Phys. Sin, 2013, 62(7):
.
doi:10.7498/aps.62.076101. |
|
|
|
|