20单元双压电片变形镜的性能测试与闭环校正实验研究

CSTR: 32037.14.aps.58.4717

Performance test and closed-loop correction experiment of a 20-element bimorph deformable mirror

CSTR: 32037.14.aps.58.4717
PDF
导出引用
  • 测试了20单元双压电片变形镜的影响函数、变形量-电压特性曲线、面形热稳定性和频率响应特性,并搭建基于20单元双压电片变形镜和13×13阵列哈特曼传感器的自适应光学系统,测试了该变形镜对自身初始面形和静态像差板的闭环校正效果.实验结果表明,各分立电极变形量对加载电压的平均灵敏度为0.57 μm/100 V,离焦电极的灵敏度为1.09 μm/100 V;各电极变形量-电压响应曲线均存在迟滞现象,滞后量在最大变形量的10%以内;在温度引起的双压电片变形镜面形变化中离焦项贡献最大,其变化对温度也最为敏感,在13℃

     

    Performances of a 20-element bimorph deformable mirror are measured and a closed-loop adaptive optics system, which consists of the bimorph mirror and a Shack-Hartmann wave-front sensor with 13×13 sub-aperture array, is established to test the mirror’s capability of correcting static aberrations. Experimental results show that, the mean displacement response to voltage of discrete actuators is 0.57 μm/100 V and that of defocus electrode is 1.09 μm/100 V. The hysteresis of actuator is limited to ten percents of the maximal displacement. With the ambient temperature increasing from 13℃ to 25℃, the P-V of defocus of the mirror surface is enlarged by 1.37 μm and the P-V of other distortions, such as astigmatism, coma and spherical aberration, is enlarged by 0.15 μm. The frequency response of the bimorph mirror is about 4 kHz. When the closed-loop control is on, intensity distributions at the image plane of distorted wave-fronts are effectively improved. The Strehl ratios increase from 0.02 to 0.95, and 0.03 to 0.54, respectively.

     