基于蒙特卡罗方法的NaI探测器效率刻度及其测量轫致辐射实验

CSTR: 32037.14.aps.63.180702

Efficiency calibration for a NaI scintillation detector based on Monte-Carlo process and preliminary measurements of bremsstrahlung

CSTR: 32037.14.aps.63.180702
PDF
导出引用
  • 为了将NaI探测器更好地应用到轫致辐射谱测量工作中,对一套NaI探测器做了研究:利用137Cs,60Co等同位素γ源,结合蒙特卡罗方法,得到全能峰效率的模拟值与实验测量值符合得较好;利用蒙特卡罗N粒子编码模拟NaI对不同能量光子的响应,得到了该探测器对光子的能量响应,并将获得的能量响应用于轫致辐射的解谱工作,解谱结果与原始谱符合得很好;将该探测器应用到强流电子束打靶轫致辐射测量实验中,对轫致辐射在NaI探测器中的响应做了初步测量.

     

    In order to better apply the NaI scintillation spectrometer to bremsstrahlung measurements, the energy response function of a NaI detector spectrometer system is studied by using 137Cs and 60Co sources based on Monte Carlo N particle transport code (MCNP) process. Simulated and measured almighty peak efficiency are in good agreement. An energy response matrix (ERM) is obtained by simulating photons with a certain energy incident on the NaI crystal in MCNP process, through deconvoluting the detected spectrum of NaI using the ERM, and the results of the deconvolution accord well with those from the original spectrum. Furthermore, the NaI detector is used to preliminarily detect its response to bremsstrahlung generated by high intensity electrons bombarding a target of 1.5 mm thickness.

     