基于FPGA实时非线性校正的共振扫描激光共聚焦显微成像

Resonant-Scanning Laser Confocal Microscopy (RS-LCM) with FPGA-Based Real-Time Nonlinear Correction

PDF
导出引用
  • 共聚焦显微技术具备优异的离焦背景抑制能力,是三维显微成像的主流技术。但其依赖线性振镜逐点扫描,成像速度慢。共振扫描是实现高速扫描的有效途径。针对高速共聚焦成像及共振扫描非线性畸变问题,本文构建了一种基于FPGA实时非线性校正的共振扫描激光共聚焦显微系统(RS-LCM)。该技术利用一共振和线性振镜组合实现对焦点的二维扫描,并利用FPGA同步控制激光调制、振镜扫描与单光子探测器采集。建立时间—空间位置映射模型,通过插值重采样实现图像实时畸变校正,克服了共振振镜非线性运动引起的空间采样不均匀问题。与传统共聚焦显微技术相比,该技术成像速度提升了20倍;相较于现有的共振扫描图像重建技术,基于FPGA的板内重建速度提升达115倍。此外,将RS-LCM与单颗粒追踪相结合,实现对动态微纳颗粒二维扩散过程的跟踪观测,并对其扩散系数进行了定量测量。

     

    Laser scanning confocal microscopy (LSCM) provides effective suppression of out-of-focus background and high-contrast optical sectioning, but its imaging speed is limited by point-by-point scanning with conventional galvanometric mirrors. To improve the imaging speed of confocal microscopy, and address the nonlinear distortion inherent in resonant scanning, we developed a resonant-scanning laser confocal microscopy (RS-LCM) system with FPGA-based real-time nonlinear correction. In RS-LCM, a 12 kHz/line resonant scanner was employed for high-speed scanning along the x-direction, while a linear galvanometric mirror performed line-by-line scanning along the y-direction. FPGA was used to synchronously control laser modulation, scanner motion, and single-photon detection. To compensate for the intrinsic nonlinear spatiotemporal distortion of resonant scanning, dedicated FPGA hardware logic was developed to establish an accurate spatiotemporal mapping relationship through parallel real-time computation, and interpolation-based resampling was employed to achieve millisecond-level online nonlinear correction. Experimental results show that RS-LCM increases the imaging speed by approximately 20-fold compared with conventional point-scanning LSCM while maintaining comparable lateral resolution. RS-LCM was further combined with single-particle tracking (SPT) to investigate the dynamics of rapidly moving particles. In SPT analysis, the diffusion coefficient of each particle was obtained by fitting the mean-square displacement (MSD) of each valid trajectory, enabling characterization of particle diffusion in samples with heterogeneous particle-size distributions. Owing to its high imaging speed, RS-LCM effectively reduces motion blur caused by rapid Brownian motion and enables reliable extraction of particle trajectories and dynamic parameters. For particles dispersed in a 39 wt% glycerol–water solution, the diffusion coefficient measured by SPT was 0.77 μm2·s-1, which agrees well with the value of 0.79 μm2·s-1 obtained by fluorescence correlation spectroscopy (FCS). These results demonstrate that resonant scanning not only improves the imaging speed of confocal microscopy, but also extends its capability for reliable tracking and quantitative measurement of rapidly diffusing micro- and nanoparticles. Future work will focus on further improving photon-collection efficiency under high-speed scanning conditions and extending the technique to more complex dynamic biological samples.

     

    目录

    /

    返回文章
    返回