同步辐射软X射线源用于软X射线探测元件定标

CSTR: 32037.14.aps.46.650

CALIBRASION OF THE ENERGY RESPONSE FOR THE SOFT X-RAY DETECTIONS ELEMENTS WITH THE BEIJING- SYNCHROTRON RADIATION FACILITY

CSTR: 32037.14.aps.46.650
PDF
导出引用
  • 用于惯性约束聚变诊断的软X射线探测元器件的能量响应曲线定标,利用北京同步辐射装置(BSRF-3B1束线,束流为20—80mA,光子能量为250—1000eV,通量约为1012photon/s·mm2·mr2·0.1%band width)及反射率计靶室,采用AXUV-100硅光二极管作源强绝对监测.对X射线二极管(XRD)及金刚石光电导探测器等五种探测元件进行能量响应曲线定标,获得初步实验结果,并对数据进行了分析

     

    Calibration of the energy response for the soft X-ray detection elements using in the ICF diagnostic are reported.The Beijing synchrotron Radiation Facility (3B1 beam line,photon energy~250—1000eV,radiation flux ~1012photon/s·mm2·mr2·0.1% band-width,beam current ~20—80mA) and target chamber of the reflectometer are used.The source intensity is monitored by AXUV-100 photodiode.The calibration of the energy response for the XRD,filter light film,Ⅱa type diamond photoconductive detector,transmission grating and soft X-ray film are given and analysed.

     