基于强耦合混合惰性气体的量子精密测量理论与实验研究

Theoretical and Experimental Study on Quantum Sensing Based on Strongly Coupled Hybrid Noble Gases

PDF
导出引用
  • 弱磁探测是量子精密测量领域的重要研究方向之一。基于不同物理体系的量子弱磁传感器已实现飞特斯拉乃至阿特斯拉量级的灵敏度,有力推动了精密测量物理及其相关交叉学科的发展。然而,当量子弱磁传感器的测量频率拓展至亚赫兹及更低频段时,系统性能将受到仪器固有 1/f 噪声及长期漂移等低频噪声的严重制约,导致灵敏度大幅降低。为抑制此类低频噪声,本文发展了基于 129Xe-131Xe 双惰性气体混合核自旋体系的量子传感器,重点从理论上分析了混合核自旋体系对多种噪声的响应特性。依据我们的理论分析,设计并实验评估了激光功率噪声与频率噪声、气室温度、偏置磁场等主要噪声来源,并相应发展了实验上的噪声抑制技术。在此基础上,从理论与实验两方面系统研究了混合惰性气体核自旋与反馈线圈的相互作用。结果表明,反馈线圈可实现混合惰性气体核自旋之间的强耦合,且在此条件下体系将呈现非厄米例外点及其对应的响应增强现象。实验结果显示,采用特定设计的反馈线圈可将量子弱磁传感器的磁场灵敏度提升至少 3.5倍。我们将该自旋体系应用于暗物质等超越标准模型新物理所产生的赝磁场探测,实验展示了该传感器对暗物质赝磁场的探测能力,在 0.001–0.1 Hz 频段内灵敏度均优于 43 fT/Hz1/2。本文所提出的弱磁探测新方法有望推广至其他混合惰性气体,预期实现更高的探测灵敏度,同时也可为暗物质、奇异相互作用等基础物理研究提供新型探测手段。

     

    Sensitive detection of weak magnetic fields is an important application of quantum sensing, with implications for fundamental physics, inertial navigation, and biomagnetic measurements. Although spinbased sensors have achieved high sensitivities over a broad frequency range, extending their performance to sub-hertz frequencies remains challenging because slow fluctuations of experimental parameters and longterm drifts can dominate the measurement noise. Here, we theoretically investigate and experimentally demonstrate a quantum sensor based on a hybrid nuclear-spin system of 129Xe and 131Xe, combining controllable feedback-mediated coupling with systematic suppression of low-frequency technical noise. The two xenon isotopes are polarized through spin-exchange collisions with optically pumped 87Rb atoms, which also provide an optical readout of their transverse spin precession. Their precession frequencies are continuously tracked using phase-locked feedback, and a differential combination of the two frequency signals suppresses common magnetic-field fluctuations while retaining sensitivity to pseudomagnetic perturbations. We establish a dynamical model of the hybrid spin system and analyze how fluctuations of the optical and magnetic parameters enter the measured output. Guided by this analysis, we experimentally characterize the contributions from pump and probe laser power and frequency fluctuations, vapor-cell temperature variations, and bias-field fluctuations. Optimization of the operating phase reduces the coupling of pump-power fluctuations to the differential signal, while control of the remaining experimental parameters improves the stability required for low-frequency measurements. These measurements identify the dominant technical noise sources and provide a basis for evaluating the sensor performance.
    To enhance the signal response, we introduce an external feedback coil that establishes a tunable effective interaction between the two nuclear-spin ensembles. This interaction modifies the coupled spin dynamics and produces a non-Hermitian exceptional point, near which the difference between the two mode frequencies exhibits an enhanced response to weak perturbations. Under the experimental operating conditions, the response slope to the target pseudomagnetic field is increased by approximately a factor of 3.5. We distinguish this enhancement of signal responsivity from an improvement in the fundamental quantum limit of parameter estimation. The present sensor operates in a regime limited by technical noise, and its practical sensitivity is determined jointly by the signal response and the measured output noise. The performance improvement therefore relies on combining response enhancement with noise suppression.
    The resulting sensor achieves a pseudomagnetic-field sensitivity better than 43 fT/Hz1/2 throughout the 0.001–0.1 Hz frequency range. These results demonstrate a route to sensitive low-frequency measurements using coupled noble-gas nuclear spins. The combination of differential spin readout, optimized operating conditions, and controllable feedback-mediated interactions provides a platform for further im-

     

    目录

    /

    返回文章
    返回