Laser scanning confocal microscopy (LSCM) provides effective suppression of out-of-focus background and high-contrast optical sectioning, but its imaging speed is limited by point-by-point scanning with conventional galvanometric mirrors. To improve the imaging speed of confocal microscopy, and address the nonlinear distortion inherent in resonant scanning, we developed a resonant-scanning laser confocal microscopy (RS-LCM) system with FPGA-based real-time nonlinear correction. In RS-LCM, a 12 kHz/line resonant scanner was employed for high-speed scanning along the x-direction, while a linear galvanometric mirror performed line-by-line scanning along the y-direction. FPGA was used to synchronously control laser modulation, scanner motion, and single-photon detection. To compensate for the intrinsic nonlinear spatiotemporal distortion of resonant scanning, dedicated FPGA hardware logic was developed to establish an accurate spatiotemporal mapping relationship through parallel real-time computation, and interpolation-based resampling was employed to achieve millisecond-level online nonlinear correction. Experimental results show that RS-LCM increases the imaging speed by approximately 20-fold compared with conventional point-scanning LSCM while maintaining comparable lateral resolution. RS-LCM was further combined with single-particle tracking (SPT) to investigate the dynamics of rapidly moving particles. In SPT analysis, the diffusion coefficient of each particle was obtained by fitting the mean-square displacement (MSD) of each valid trajectory, enabling characterization of particle diffusion in samples with heterogeneous particle-size distributions. Owing to its high imaging speed, RS-LCM effectively reduces motion blur caused by rapid Brownian motion and enables reliable extraction of particle trajectories and dynamic parameters. For particles dispersed in a 39 wt% glycerol–water solution, the diffusion coefficient measured by SPT was 0.77 μm
2·s
-1, which agrees well with the value of 0.79 μm
2·s
-1 obtained by fluorescence correlation spectroscopy (FCS). These results demonstrate that resonant scanning not only improves the imaging speed of confocal microscopy, but also extends its capability for reliable tracking and quantitative measurement of rapidly diffusing micro- and nanoparticles. Future work will focus on further improving photon-collection efficiency under high-speed scanning conditions and extending the technique to more complex dynamic biological samples.