We report the development, systematic evaluation, and extended absolute-frequency measurement of the NTSC-Sr2 strontium optical lattice clock at the National Time Service Center. Laser-cooled
87Sr atoms are transported by a movable one-dimensional optical lattice over 17.29 mm into a copper blackbody-radiation shield stabilized near 273.15 K. Clock spectroscopy is performed at a lattice depth of 39.8
ER. A 400 ms Rabi interrogation produces a 2.5 Hz linewidth, close to the Fourier limit. Interleaved locking of two digital servos to the same atomic transition gives an equivalent single-servo fractional-frequency instability of 3.6 × 10
-16(
τ/
s)
-1/2. Because the two servos share the atomic preparation, trapping and detection sequence, as well as the clock laser, this result characterizes the short-term servo and detection performance rather than an independent comparison between two clocks. The evaluated fractional systematic correction is −3707.5 × 10
-18 with a total uncertainty of 2.0 × 10
-18; the dominant contribution is the lattice-light Stark shift, with an uncertainty of 1.6 × 10
-18, followed by the blackbody-radiation contribution of 0.7 × 10
-18. For the absolute-frequency measurement, NTSC-Sr2 is linked to the SI second through HM085, UTC(NTSC), and International Atomic Time. Three additional Circular T intervals (461, 462, and 464) extend the previously reported one-year campaign to 15 measurement months, with a mean uptime of 28.3% and approximately 3.1 × 103 hour of valid data. The weighted frequency of the
87Sr
1S
0–
3P
0 transition is 429 228 004 229 872.97(6) Hz, corresponding to a relative uncertainty of 1.5 × 10
-16. These results establish an end-to-end capability encompassing clock operation, systematic-shift evaluation, and SI-traceable absolute-frequency measurement, and provide an independent data set for optical-clock contributions to international time-scale calibration and the future redefinition of the second.