Cardiovascular disease, especially hypertension, has become one of the biggest killers of human life and health. Therefore, early detection and prevention of cardiovascular diseases are of great significance for people’s health. In this paper, we explore the relationship between aortic valve heart sound signal and blood pressure, and propose a method of non-invasively estimating blood pressure based on heart sound signals. First, according to the relationship between blood pressure and the heart sound signal, both the time interval between the peak point of the first heart sound and that of the second heart sound and the kurtosis of the second heart sound are extracted as features. Then the time interval between the first heart sound and the second heart sound, and the kurtosis of the second heart sound are linearly fitted to the measured blood pressure. Finally, according to the linear relationship between heart sound and blood pressure, a blood pressure evaluation formula based on the heart sound is established. The experimental results show that the time interval between the peak point of the first heart sound and that of the second heart sound, and the kurtosis of the second heart sound can be used as the characteristic parameters of blood pressure evaluation, which have a good linear relationship with blood pressure. The goodness of fit is 0.801 and 0.765, separately. The average error between the blood pressure calculated from the blood pressure calculation formula and the blood pressure measured by a commercial electronic sphygmomanometer is less than 5 mmHg, and the standard deviation is less than 8 mmHg. For hypertensive patients, the time interval between the peak point of the first heart sound and the second heart sound is shortened, and the kurtosis of the second heart sound is increased, which is a typical feature of heart sounds in patients with hypertension. Compared with the traditional blood pressure calculation method, the blood pressure assessment method proposed in this paper only needs to collect heart sound signals to effectively assess the blood pressure. The method is convenient to operate and can be used for continuously monitoring the blood pressure, and is especially suitable for monitoring the blood pressures of infants, disabled patients with limbs, and disabilities in certain medical environments.
-
Keywords:
- blood pressure /
- aortic valve heart sound /
- peak detection /
- kurtosis
[1] WHO NCD Mortality and Morbidity http://www.who.int/gho/ncd/mortality_morbidity/en/ [2015−09−07]
[2] Durand L G, Pibarot P 1995 Crit. Rev. Biomed. Eng. 23 163Google Scholar
[3] Liu C C, Springer D B, Clifford G D 2017 Physiol. Meas. 38 1730Google Scholar
[4] Bartels A, Harder D 1992 Clin. Phys. Physiol. Meas. 13 249Google Scholar
[5] Tanigawa N, Smith D, Craige E 1991 Jpn. Circ. J. 55 737Google Scholar
[6] Zhang X Y, Zhang Y T 2006 Proceedings of the IEEE Engineering in Medicine and Biology Society New York, USA, August 30−September 3, 2006 p2888
[7] Peng R C, Yan W R, Zhang N L, Lin W H, Zhou X L, Zhang Y T 2015 Sensors-basel 15 23653Google Scholar
[8] Bombardini T, Gemignani V, Bianchini E, Venneri L, Petersen C 2008 Cardiovasc. Ultrasoun. 6 41Google Scholar
[9] Tao Y W CN104510491A [2015−04−15]
[10] 黄正钦 2014 硕士学位论文 (昆明: 云南大学)
Huang Z Q 2014 M. S. Thesis (Kunming: Yunnan University) (in Chinese)
[11] Shriram R, Wakankar A, Daimiwal N 2010 Proceedings of International Conference on Bioinformatics and Biomedical Technology Chengdu, China, April 16−18, 2010 p100971242
[12] Landowne M 1957 Circ. Res. 5 594Google Scholar
[13] Payne R A, Symeonides C N, Webb D J, Maxwell S R J 2006 J. Appl. Physiol. 100 136Google Scholar
[14] 蔡任圃 2012 科学月刊 507 236
Cai R P 2012 Science Monthly 507 236
[15] 周静, 杨永明, 何为 2005 中国生物医学工程学报 24 685Google Scholar
Zhou J, Yang Y M, He W 2005 Chin. J. Biomed. Eng. 24 685Google Scholar
[16] Ding X R, Zhao R, Yang G Z, Pettigrew R 2016 IEEE J. Biomed. Health 20 1Google Scholar
[17] Lim K H, Shin Y D, Park S H, Bae J H, Lee H J, Kim S J 2013 Pak. J. Med. Sci. 29 1023
-
表 1 编号5-2测试者第二心音峰值点与脉搏波峰值点在时间上的对应关系
Table 1. Correspondence between the peak point of the second heart sound and pulse wave of No. 5-2 subject.
信号类型 时间/ms 第二心音
峰值点205 976 1722 2481 3233 3956 4697 5446 6184 脉搏波
峰值点209 982 1727 2486 3238 3960 4704 5454 6189 表 2 HSTT与血压拟合优度
Table 2. Goodness of fit between HSTT and BP.
状态 SBP DBP 均值 静止 0.847 0.774 0.801 运动 0.826 0.753 运动后恢复 0.839 0.768 表 3 第二心音峭度(K )与血压拟合优度
Table 3. The second heart sound kurtosis (K ) and goodness of fit of blood pressure.
状态 SBP DBP 均值 静止 0.824 0.703 0.765 运动 0.818 0.724 运动后恢复 0.806 0.718 表 4 第5组测试者3种状态下心音与血压相关数据统计结果(均值)
Table 4. Data related to heart sounds and blood pressure in the 5th group.
测试者状态 HSTT/ms K 计算${\rm{SBP}}$/mmHg 实测${\rm{SBP}}$/mmHg 计算${\rm{DBP}}$/mmHg 实测${\rm{DBP}}$/mmHg 静止 277.9 3.31 120.28 122.6 71.86 71.2 运动 246.7 4.19 139.00 140.2 84.34 83.0 运动后恢复 264.3 3.72 128.43 128.8 77.29 75.8 表 5 所有测试者数据参数统计结果
Table 5. Data parameter statistics of all sujects.
参数 最大值 最小值 平均值 ${{\rm{CC}}_{\rm{SBP}}}$ 0.923 0.542 0.781 ${{\rm{CC}}_{\rm{DBP}}}$ 0.891 0.515 0.748 ${{\rm{MAE}}_{\rm{SBP}}}$/mmHg 6.245 1.054 2.712 ${{\rm{MAE}}_{\rm{DBP}}}$/mmHg 5.968 1.628 3.264 ${{\rm{ME}}_{\rm{SBP}}}$/mmHg 1.091 –1.331 –0.401 ${{\rm{ME}}_{\rm{DBP}}}$/mmHg 0.526 –1.254 –0.812 ${{\rm{SD}}_{\rm{SBP}}}$/mmHg 8.534 2.723 3.584 ${{\rm{SD}}_{\rm{DBP}}}$/mmHg 6.234 2.132 4.053 表 6 15名测试者血压和心音相关数据统计结果
Table 6. Datas related to heart sounds and blood pressure of 15 testers.
测试者状态 ${\rm{HSTT}}$/ms K 计算SBP/mmHg 实测SBP/mmHg 计算DBP/mmHg 实测DBP/mmHg 健康 279.4 3.38 122.46 121.63 73.16 72.02 高血压患者 267.8 4.52 151.94 148.58 84.34 81.17 -
[1] WHO NCD Mortality and Morbidity http://www.who.int/gho/ncd/mortality_morbidity/en/ [2015−09−07]
[2] Durand L G, Pibarot P 1995 Crit. Rev. Biomed. Eng. 23 163Google Scholar
[3] Liu C C, Springer D B, Clifford G D 2017 Physiol. Meas. 38 1730Google Scholar
[4] Bartels A, Harder D 1992 Clin. Phys. Physiol. Meas. 13 249Google Scholar
[5] Tanigawa N, Smith D, Craige E 1991 Jpn. Circ. J. 55 737Google Scholar
[6] Zhang X Y, Zhang Y T 2006 Proceedings of the IEEE Engineering in Medicine and Biology Society New York, USA, August 30−September 3, 2006 p2888
[7] Peng R C, Yan W R, Zhang N L, Lin W H, Zhou X L, Zhang Y T 2015 Sensors-basel 15 23653Google Scholar
[8] Bombardini T, Gemignani V, Bianchini E, Venneri L, Petersen C 2008 Cardiovasc. Ultrasoun. 6 41Google Scholar
[9] Tao Y W CN104510491A [2015−04−15]
[10] 黄正钦 2014 硕士学位论文 (昆明: 云南大学)
Huang Z Q 2014 M. S. Thesis (Kunming: Yunnan University) (in Chinese)
[11] Shriram R, Wakankar A, Daimiwal N 2010 Proceedings of International Conference on Bioinformatics and Biomedical Technology Chengdu, China, April 16−18, 2010 p100971242
[12] Landowne M 1957 Circ. Res. 5 594Google Scholar
[13] Payne R A, Symeonides C N, Webb D J, Maxwell S R J 2006 J. Appl. Physiol. 100 136Google Scholar
[14] 蔡任圃 2012 科学月刊 507 236
Cai R P 2012 Science Monthly 507 236
[15] 周静, 杨永明, 何为 2005 中国生物医学工程学报 24 685Google Scholar
Zhou J, Yang Y M, He W 2005 Chin. J. Biomed. Eng. 24 685Google Scholar
[16] Ding X R, Zhao R, Yang G Z, Pettigrew R 2016 IEEE J. Biomed. Health 20 1Google Scholar
[17] Lim K H, Shin Y D, Park S H, Bae J H, Lee H J, Kim S J 2013 Pak. J. Med. Sci. 29 1023
Catalog
Metrics
- Abstract views: 11601
- PDF Downloads: 138
- Cited By: 0