-
在扫频光学相干层析系统中, 远心扫描模式造成角膜图像中存在伪影、部分结构缺失及低信噪比区域, 影响了角膜轮廓提取的精度. 针对该问题, 本文提出了一种针对低质量角膜图像的轮廓自动提取算法. 该算法首先依据图像标准差分布将图像划分为高、低信噪比区域; 针对高信噪比区域, 通过峰值点定位法获取角膜轮廓; 针对低信噪比区域, 通过连续帧图像间配准叠加实现图像增强, 为低信噪比区域提供参考轮廓点, 再通过权衡参考轮廓点与局部直线拟合结果的优劣, 实现角膜轮廓定位; 最后, 通过全局多项式拟合实现对全区域的角膜整体轮廓信息. 对光学眼模型进行实验, 结果表明, 与已有算法相比, 本文算法对角膜轮廓的提取精度平均提高了4.9%.In a swept source-optical coherence tomography system, the telecentric scanning mode gives rise to central saturation artifacts,partial structural loss, and low SNR (signal-to-noise ratio) area in the corneal image, which affects the accuracy of corneal contour extraction. In order to solve this problem, in this paper we propose an automatic extraction algorithm for corneal image of low quality. This algorithm divides the image into high and low SNR region according to the standard deviation distribution of the cornea image. For the high SNR region, we localize the peak point to extract the contour. For the low SNR region, image enhancement is achieved by the registration and superposition of successive frames, which provides reference contour points for low SNR areas. Then corneal contour localization is achieved by weighing the advantages and disadvantages of reference contour points and local line fitting results. Finally, global polynomial fitting is used to achieve the whole corneal contour information. Experiments on the optical eye model show that comparing with the existing algorithms, the accuracy of corneal contour extraction is improved by 4.9% on average.
-
Keywords:
- optical coherence tomography /
- swept-source optical coherence tomography /
- cornea /
- contour extraction
[1] Wang J F, Zheng W, Lin K, Huang Z W 2016 Opt. Lett. 41 3045Google Scholar
[2] Kumar A, Baumann B, Hafner J, Ginner L, Augustin M, Salas M, Pircher M, Leitgeb R, Prager S, Schmidterfurth U 2016 Biomed. Opt. Express 8 207
[3] Grulkowski I, Liu J J, Potsaid B, Jayaraman V, Jiang J, Fujimoto J G, Cable A E 2013 Opt. Lett. 38 673Google Scholar
[4] Polans J, Cunefare D, Cole E, Keller B, Mettu P S, Cousins S W, Allingham M J, Izatt J A, Farsiu S 2017 Opt. Lett. 42 17Google Scholar
[5] Camino A, Jia Y L, Liu G J, Wang J, Huang D 2017 Biomed. Opt. Express 8 3053Google Scholar
[6] Kaluzny B J, Karnowski K, Szkulmowski M, Wojtkowski M, Gora M 2011 Biomed. Opt. Express 2 2709Google Scholar
[7] Lawman S, Dong Y, Williams B M, Romano V, Kaye S, Harding S P, Willoughby C, Shen Y C, Zheng Y L 2016 Opt. Express 24 12395Google Scholar
[8] Lawman S, Madden P W, Vito R, Romano V, Dong Y, Mason S, Williams B M, Kaye S B, Willoughby C E, Harding S P, Shen Y C 2017 Biomed. Opt. Express 8 5579Google Scholar
[9] Wang L L, Xiong Q Z, Ge X, Bo E, Xie J, Liu X Y, Yu X J, Wang X H, Wang N S, Chen S, Wu X, Liu L B 2019 Opt. Express 27 1298Google Scholar
[10] Williams D, Zheng Y L, Bao F J, Elsheikh A 2013 J. Biomed. Opt. 18 056003
[11] Li Y, Shekhar R, Huang D 2002 Proc. SPIE 4684 167Google Scholar
[12] Li Y, Shekhar R, Huang D 2006 Ophthalmology 113 792Google Scholar
[13] Shu P, Sun Y 2012 J. Innov. Opt. Heal. Sci. 5 9
[14] LaRocca F, Chiu S J, McNabb R P, Kuo A N, Izatt J A, Farsiu S 2011 Biomed. Opt. Express 2 1524Google Scholar
[15] Keller B, Draelos M, Tang G, Farsiu S, Kuo A N, Hauser K, Izatt J A 2018 Biomed. Opt. Express 9 2716Google Scholar
[16] Santos V A, Schmetterer L, Stegmann H, Pfister M, Messner A, Schmidinger G, Garhofer G, Werkmeister R M 2019 Biomed. Opt. Express 10 622Google Scholar
[17] Otsu N 2007 IEEE. Trans. Syst. Man. Cybern. 9 62
[18] Doughty M J, Zaman M L 2000 Surv. Ophthalmol. 44 367Google Scholar
[19] Gifford P, Ahmed A, Swarbrick H A 2011 Invest. Ophth. Vis. Sci. 52 3648Google Scholar
[20] Atchison D A, Jones C E, Schmid K L, Nicola P, Pope J M, Strugnell W E, Riley R A 2004 Invest. Ophth. Vis. Sci. 45 3380Google Scholar
[21] Wang Q, Liu W W, Wu Y L, Ma Y, Zhao G Q 2017 Clin. Exp. Optom. 100 250
[22] Saenzfrances F, Gonzalezpastor E, Borregosanz L, Jerezfidalgo M, Martinezdelacasa J, Mendezhernandez C, Santosbueso E, Fernandezvidal A, Garciasanchez J, Garciafeijoo J 2012 J. FR. Ophtalmol. 35 333Google Scholar
[23] Read S A, Collins M J 2009 Optometry & Vision Science Official Publication of the American Academy of Optometry 86 170
[24] Zaki F, Wang Y H, Su H, Yuan X, Liu X 2017 Biomed. Opt. Express 8 2720Google Scholar
[25] 张强, 那彦, 李建军 2006 应用光学 27 4
Zhang Q, Na Y, Li J J 2006 J. Appl. Opt. 27 4
-
表 1 两种算法对角膜轮廓平均提取精度对比
Table 1. Comparison of the accuracy of two algorithmsfor contour extraction of high and low SNR regions.
评价指标 文献[14] 本文算法 精度提高值/% 平均误差值/pixels 2.8 0.9 4.9 偏差率/% 7.2 2.3 -
[1] Wang J F, Zheng W, Lin K, Huang Z W 2016 Opt. Lett. 41 3045Google Scholar
[2] Kumar A, Baumann B, Hafner J, Ginner L, Augustin M, Salas M, Pircher M, Leitgeb R, Prager S, Schmidterfurth U 2016 Biomed. Opt. Express 8 207
[3] Grulkowski I, Liu J J, Potsaid B, Jayaraman V, Jiang J, Fujimoto J G, Cable A E 2013 Opt. Lett. 38 673Google Scholar
[4] Polans J, Cunefare D, Cole E, Keller B, Mettu P S, Cousins S W, Allingham M J, Izatt J A, Farsiu S 2017 Opt. Lett. 42 17Google Scholar
[5] Camino A, Jia Y L, Liu G J, Wang J, Huang D 2017 Biomed. Opt. Express 8 3053Google Scholar
[6] Kaluzny B J, Karnowski K, Szkulmowski M, Wojtkowski M, Gora M 2011 Biomed. Opt. Express 2 2709Google Scholar
[7] Lawman S, Dong Y, Williams B M, Romano V, Kaye S, Harding S P, Willoughby C, Shen Y C, Zheng Y L 2016 Opt. Express 24 12395Google Scholar
[8] Lawman S, Madden P W, Vito R, Romano V, Dong Y, Mason S, Williams B M, Kaye S B, Willoughby C E, Harding S P, Shen Y C 2017 Biomed. Opt. Express 8 5579Google Scholar
[9] Wang L L, Xiong Q Z, Ge X, Bo E, Xie J, Liu X Y, Yu X J, Wang X H, Wang N S, Chen S, Wu X, Liu L B 2019 Opt. Express 27 1298Google Scholar
[10] Williams D, Zheng Y L, Bao F J, Elsheikh A 2013 J. Biomed. Opt. 18 056003
[11] Li Y, Shekhar R, Huang D 2002 Proc. SPIE 4684 167Google Scholar
[12] Li Y, Shekhar R, Huang D 2006 Ophthalmology 113 792Google Scholar
[13] Shu P, Sun Y 2012 J. Innov. Opt. Heal. Sci. 5 9
[14] LaRocca F, Chiu S J, McNabb R P, Kuo A N, Izatt J A, Farsiu S 2011 Biomed. Opt. Express 2 1524Google Scholar
[15] Keller B, Draelos M, Tang G, Farsiu S, Kuo A N, Hauser K, Izatt J A 2018 Biomed. Opt. Express 9 2716Google Scholar
[16] Santos V A, Schmetterer L, Stegmann H, Pfister M, Messner A, Schmidinger G, Garhofer G, Werkmeister R M 2019 Biomed. Opt. Express 10 622Google Scholar
[17] Otsu N 2007 IEEE. Trans. Syst. Man. Cybern. 9 62
[18] Doughty M J, Zaman M L 2000 Surv. Ophthalmol. 44 367Google Scholar
[19] Gifford P, Ahmed A, Swarbrick H A 2011 Invest. Ophth. Vis. Sci. 52 3648Google Scholar
[20] Atchison D A, Jones C E, Schmid K L, Nicola P, Pope J M, Strugnell W E, Riley R A 2004 Invest. Ophth. Vis. Sci. 45 3380Google Scholar
[21] Wang Q, Liu W W, Wu Y L, Ma Y, Zhao G Q 2017 Clin. Exp. Optom. 100 250
[22] Saenzfrances F, Gonzalezpastor E, Borregosanz L, Jerezfidalgo M, Martinezdelacasa J, Mendezhernandez C, Santosbueso E, Fernandezvidal A, Garciasanchez J, Garciafeijoo J 2012 J. FR. Ophtalmol. 35 333Google Scholar
[23] Read S A, Collins M J 2009 Optometry & Vision Science Official Publication of the American Academy of Optometry 86 170
[24] Zaki F, Wang Y H, Su H, Yuan X, Liu X 2017 Biomed. Opt. Express 8 2720Google Scholar
[25] 张强, 那彦, 李建军 2006 应用光学 27 4
Zhang Q, Na Y, Li J J 2006 J. Appl. Opt. 27 4
计量
- 文章访问数: 7353
- PDF下载量: 64
- 被引次数: 0