单分子定位显微(single molecule localization microscopy, SMLM)成像技术利用荧光分子的稀疏发光、探测及定位, 实现了纳米级空间分辨率的超分辨成像. 为了提高其时间分辨率, 需要提高同时发光的荧光分子密度. 但随着分子密度的提高, 不同分子的点扩散函数(point spread function, PSF)在探测器上将发生严重的重叠现象, 导致空间分辨率降低, 尤其是在进行三维SMLM成像时. 为了解决这一问题, 本文提出了一种基于正交像散的高密度三维单分子定位超分辨成像方法, 并对该方法进行分析和数值模拟研究. 该方法的核心是在单分子定位显微镜中将采集的荧光分成两束成像在同一个探测器的两个区域, 并在两个通道中各引入一个光学参数相同但取向相互正交的柱透镜, 实现对同一个荧光分子正负两个像散PSF图像的同时探测, 然后建立该成像过程的线性投影模型, 利用压缩感知算法求解出荧光分子的三维定位信息. 结果表明, 由于两个正交柱透镜产生的一组正交像散PSF对作为一个分子的系统响应时具有较低的相关性, 该方法的高密度三维定位准确性可显著优于采用单个柱透镜的传统像散方法, 且离焦程度越大两个像散PSF的形状差异越大, 这种准确定位的优势就越明显.Single molecule localization microscopy (SMLM) detects and locates sparsely luminous single fluorescent molecules to achieve super-resolution imaging at nanoscale spatial resolution. In order to improve the temporal resolution, it is necessary to increase the density of the simultaneously emitting molecules. However, with the increase of the density, the point spread function (PSF) of different molecules will overlap severely on the detector, resulting in reduced spatial resolution, especially for three-dimensional (3D) SMLM. To solve this problem, a high density 3D-SMLM imaging method based on orthogonal astigmatism is proposed. Analysis and numerical simulation study for the method are carried out and presented. The main idea of the proposed orthogonal astigmatic method is to split the collected fluorescence in a SMLM microscope into two beams, each of which passes through a separate channel with a cylindrical lens and arrives at a specific region on the same detector. The two cylindrical lenses have the same optical parameters, but their orientations are set to be orthogonal to each other. They are used to obtain both positive and negative astigmatic PSF images of the same fluorescent molecule. Then, a linear projection model of the imaging process is established, and the 3D localization of the fluorescent molecules is realized by using a compression sensing algorithm. The results show that the two orthogonal cylindrical lenses produce a pair of astigmatic PSFs for one single molecule so that different PSF pairs between different molecules have lower mutual correlation, and thus the 3D localization accuracy for high density imaging can be significantly improved as compared with traditional astigmatic method, in which one single cylindrical lens is used. The larger the defocusing degree, the greater the shape difference between the two astigmatic PSFs is, and the more obvious this advantage.
-
Keywords:
- single molecule localization microscopy /
- orthogonal astigmatism /
- three-dimensional single molecule localization /
- high-density imaging
[1] Rust M J, Bates M, Zhuang X W 2006 Nat. Methods 3 793Google Scholar
[2] Bates M, Jones S A, Zhuang X 2013 Cold Spring Harbor Protoc. 2013 498Google Scholar
[3] Huang B, Wang W, Bates M, Zhuang X 2008 Science 319 810Google Scholar
[4] Huang B, Jones S A, Brandenburg B, Zhuang X W 2008 Nat. Methods 5 1047Google Scholar
[5] Betzig E, Patterson G H, Sougrat R, Lindwasser O W, Olenych S, Bonifacino J S, Davidson M W, Lippincott-Schwartz J, Hess H F 2006 Science 313 1642Google Scholar
[6] Holden S J, Uphoff S, Kapanidis A N 2011 Nat. Methods 8 279Google Scholar
[7] Zhu L, Zhang W, Elnatan D, Huang B 2012 Nat. Methods 9 721Google Scholar
[8] Babcock H, Sigal Y M, Zhuang X 2012 Opt. Nanoscopy 1 6Google Scholar
[9] Gu L, Sheng Y, Chen Y, Chang H, Zhang Y, Lv P, Ji W, Xu T 2014 Biophys. J. 106 2443Google Scholar
[10] Min J, Holden S J, Carlini L, Unser M, Manley S, Ye J C 2014 Biomedical Optics Express 5 3935Google Scholar
[11] Huang J Q, Sun M Z, Gumpper K, Chi Y J, Ma J J 2015 Biomed. Opt. Express 6 902Google Scholar
[12] Huang J Q, Sun M Z, Ma J J, Chi Y J 2017 IEEE Transa. Comput. Imaging 3 763Google Scholar
[13] Von Middendorff C, Egner A, Geisler C, Hell S, Schonle A 2008 Opt. Express 16 20774Google Scholar
-
图 1 正交像散单分子定位成像光路和原理示意图 (a)成像光路示意图; (b)正交像散PSF图像对; (c)正交像散校准曲线. OL, 物镜; DM, 二向色镜; EF, 发射滤光片; BS, 分束器; A, 光阑; CL, 柱透镜; L, 透镜; M, 平面镜; EMCCD, 电子倍增电荷耦合器件
Fig. 1. Schematic diagram of the optical path and principle of single molecule localization imaging based on orthogonal astigmatism: (a) Optical path; (b) orthogonal astigmatic PSFs; (c) calibration curves. OL, objective lens; DM, dichroic mirror; EF, emission filter; BS, beam splitter; A, aperture; CL, cylindrical lens; L, lens; M, mirror; EMCCD, electron-multiplying charge-coupled device.
图 7 平行线的模拟成像结果 (a), (c), (e)正交像散法; (b), (d), (f)传统像散法; (c)—(f)间距最小(50 nm)平行线的放大图; (g)图(c)和(d)的绿色方框区域的强度分布曲线; (h)图(e)和(f)的黄色方框区域的强度分布曲线. 标尺大小: (a), (b) 500 nm; (c)—(f) 200 nm
Fig. 7. Simulated imaging results of parallel line structures: (a), (c), (e) Orthogonal astigmatic method; (b), (d), (f) traditional astigmatic method; (c)–(f) zoomed-in view of the minimum spacing (50 nm) lines; (g) cross-sectional profiles of the green boxes in panel (c) and (d); (h) cross-sectional profiles of the yellow boxes in panel (e) and (f). Scale bars: (a), (b) 500 nm; (c)–(f) 200 nm.
图 8 双通道图像偏差的影响和有无图像偏差及配准的模拟成像结果比较 (a)定位准确性随横向偏移量、旋转角度和缩放倍率的变化; (b)有偏差双通道图像配准后的模拟成像结果; (c)无偏差双通道图像的模拟成像结果. 标尺大小: 500 nm
Fig. 8. Influence of deviation between two channel images, and comparison of simulated imaging results with and without image deviation and registration: (a) Localization accuracy versus lateral offset, rotation angle and scaling ratio; (b) simulated image obtained after registration of biased dual channel images; (c) simulated image of unbiased dual channel images. Scale bars: 500 nm.
-
[1] Rust M J, Bates M, Zhuang X W 2006 Nat. Methods 3 793Google Scholar
[2] Bates M, Jones S A, Zhuang X 2013 Cold Spring Harbor Protoc. 2013 498Google Scholar
[3] Huang B, Wang W, Bates M, Zhuang X 2008 Science 319 810Google Scholar
[4] Huang B, Jones S A, Brandenburg B, Zhuang X W 2008 Nat. Methods 5 1047Google Scholar
[5] Betzig E, Patterson G H, Sougrat R, Lindwasser O W, Olenych S, Bonifacino J S, Davidson M W, Lippincott-Schwartz J, Hess H F 2006 Science 313 1642Google Scholar
[6] Holden S J, Uphoff S, Kapanidis A N 2011 Nat. Methods 8 279Google Scholar
[7] Zhu L, Zhang W, Elnatan D, Huang B 2012 Nat. Methods 9 721Google Scholar
[8] Babcock H, Sigal Y M, Zhuang X 2012 Opt. Nanoscopy 1 6Google Scholar
[9] Gu L, Sheng Y, Chen Y, Chang H, Zhang Y, Lv P, Ji W, Xu T 2014 Biophys. J. 106 2443Google Scholar
[10] Min J, Holden S J, Carlini L, Unser M, Manley S, Ye J C 2014 Biomedical Optics Express 5 3935Google Scholar
[11] Huang J Q, Sun M Z, Gumpper K, Chi Y J, Ma J J 2015 Biomed. Opt. Express 6 902Google Scholar
[12] Huang J Q, Sun M Z, Ma J J, Chi Y J 2017 IEEE Transa. Comput. Imaging 3 763Google Scholar
[13] Von Middendorff C, Egner A, Geisler C, Hell S, Schonle A 2008 Opt. Express 16 20774Google Scholar
计量
- 文章访问数: 4606
- PDF下载量: 69
- 被引次数: 0